Novel approaches to quality control and external quality assessment for platelet function testing with a focus on the platelet function analyser (PFA-100 and PFA-200)
Review Article

Novel approaches to quality control and external quality assessment for platelet function testing with a focus on the platelet function analyser (PFA-100 and PFA-200)

Emmanuel J. Favaloro1,2

1Department of Haematology, Institute of Clinical Pathology and Medical Research, NSW Health Pathology, Westmead Hospital, Westmead, NSW, Australia; 2Sydney Centres for Thrombosis and Haemostasis, Westmead, NSW, Australia

Correspondence to: Emmanuel J. Favaloro. Department of Haematology, Institute of Clinical Pathology and Medical Research, Westmead Hospital, Westmead, NSW 2145, Australia. Email: emmanuel.favaloro@health.nsw.gov.au.

Abstract: Platelet function testing is a key diagnostic activity within hematology laboratories associated with advanced hemostasis diagnostics. Platelets represent a key component of primary hemostasis, and deficiency and/or defects, either congenital or acquired, leads to bleeding diathesis in affected individuals. There are various levels of platelet function tests, from simple screens, to complex functional assays, as well as molecular analysis. Platelet function testing has evolved to now incorporate a variety of processes, such as whole blood aggregometry, light transmission aggregometry (LTA), testing by platelet function analyser (PFA) -100 (or -200), flow cytometry, and many other methodologies. Despite some of these tests being available now for decades, and/or continuously evolving, internal quality control (IQC) and external quality assessment (EQA) is limited and made difficult by the nature of the tests and test material, the latter typically representing functional cellular material (i.e., platelets). The current review looks at platelet function testing from the perspective of diagnostic screening, and highlights current limitations, as well as potential solutions that will enable more effective and accurate testing in the future. The primary focus of the review, however, is on IQC and EQA for the platelet function analyser (PFA-100 and PFA-200).

Keywords: Platelet function testing; platelet function analyser (PFA); PFA-100; PFA-200; external quality assessment (EQA); internal quality control (IQC)


Received: 18 January 2019; Accepted: 30 January 2019; Published: 31 January 2019.

doi: 10.21037/aob.2019.01.02


Introduction

Platelets represent a key component of primary hemostasis, and deficiency and/or defects in platelets, be they congenital or acquired, lead to bleeding diatheses in affected individuals (1-3). In turn, platelet function tests (PFTs), aiming to identify if platelets are functionally active or instead defective or impaired in some way, represent key diagnostic test processes within hematology laboratories associated with advanced hemostasis diagnostic facilities (4-7).

There are various levels of platelet function tests, from simple screening tests, to complex functional assays and molecular analysis (3-7). Platelet function testing has evolved to now incorporate a wide variety of test processes. These processes include (I) whole blood aggregometry (WBA), which in turn may be driven by a variety of instrumentation (3-8), (II) light transmission aggregometry (LTA) (3-7,9), representing a kind of ‘gold-standard’ in testing by nature of long historical use and wide user experience, (III) testing by automated platelet function analyser (PFA) -100 (or -200) (3-7,10), (IV) assessments by flow cytometry (3-7,11-13), and ‘(V)’ to ‘(XXVI)’ representing a large number of additional methodologies (3,6,7). The reader is referred to Table 1 for a summary of the main types of platelet function tests currently available in diagnostics. However, many other procedures may also be used to monitor anti-platelet therapies and/or in research settings. Therefore, should readers be interested in more detailed listings, they are referred to other excellent reviews (3,6,7,14).

Table 1

A summary of the types of platelet function tests currently available and used in diagnostics*

Procedure What it measures/detects Strengths/benefits/advantages Limitations/weaknesses/disadvantages
Light transmission aggregometry (LTA) Low shear platelet-to-platelet aggregation in response to a range of agonists and concentrations Gold standard. Widely used in specialized laboratories. High publication and evidence base around usage Time-consuming, complex, sample preparation, poorly standardized, and requires specialized equipment. Limited IQC and EQA
Whole blood aggregometry (WBA) Monitors changes in impedance in response to a range of agonists, sometimes including a range of agonist concentrations Simplified whole blood test, multichannel version available. Widely used in specialized laboratories although less than LTA Dependent on platelet count, older instruments require electrodes to be cleaned and recycled. Simplified system has limited utility in diagnostics, and perhaps has some utility in monitoring anti-platelet therapy. Requires specialized equipment. Not standardised. Limited IQC and no EQA available
Lumi-aggregometry Combines LTA or WBA with measurement of nucleotide release Monitors release reaction with secondary aggregation. Widely used in specialized labs, although less than LTA or WBA Semiquantitative. Requires specialized equipment. Not standardised. Limited IQC and no EQA available
PFA-100/-200 High-shear platelet adhesion and aggregation during formation of a platelet plug Whole blood test, high shear, small blood volumes, simple, rapid, POC feasibility. Very sensitive to VWD. Widely used. EQA available Inflexible; VWF, hematocrit and platelet count dependent, meaning not specific for platelet function. Requires specialized equipment. May miss some forms of mild VWD (e.g., mild type 1, alternatively called ‘low VWF as a cause of bleeding’)
Flow cytometry Measurement of platelet glycoproteins and activation markers by fluorescence Whole blood test, small blood volumes, wide variety of tests. Increasingly used in specialized labs Specialized operator, expensive, samples prone to artifact unless carefully prepared. Mainly in realm of research at the moment. Not standardised. Requires specialized equipment. Limited IQC and no EQA available

*Table is not meant to be an exhaustive list of options, but rather expresses the main procedures used in diagnostics for platelet function assessment. Many other procedures may also be used to monitor anti-platelet therapies and/or in research settings. For a detailed listing, please refer to other excellent reviews (3,6,7,14). IQC, internal quality control; EQA, external quality assessment; POC, point of care; PFA, platelet function analyser; VWD, von Willebrand disease; VWF, von Willebrand factor.

Despite many of these tests being in diagnostic use now for decades, and/or continuously evolving, internal quality control (IQC) and external quality assessment (EQA) for them is limited and made specifically difficult for diagnostic PFTs by the nature of the tests themselves, as well the test material used, with this typically representing functional cellular material (i.e., platelets). The current review therefore aims to overview platelet function testing from the perspective of diagnostic screening and highlights current limitations in IQC and EQA for PFTs, as well as highlighting potential solutions that will enable more effective and accurate testing in the future. However, the primary focus of the review is on IQC and EQA for the platelet function analyser (PFA-100 and PFA-200).


What are PFTs and what are the problems associated with these tests?

Simplistically, PFTs aim to investigate the function of platelets. Although this seems a straightforward statement of fact, how do laboratories actually achieve this, and how can they ensure the quality of PFTs?

To some extent, the answer to such questions depends on the type of tests that are performed. Platelets are complex cellular components of hemostasis (1). There are over 100 million platelets in each mL of our blood, and each platelet expresses a large number of cell surface receptors, and also houses several internal granule types that contain a myriad of (primarily pro-) haemostatic components. Simplistically, then, PFTs aim to investigate whether the platelets are working by investigation of the activity of either the cell surface receptors, or its internal components. In brief, platelets are chiefly involved in so-called primary hemostasis—meaning platelet aided formation of platelet ‘plugs’ to seal sites of vascular injury, and thereby stop bleeding (1,3,15,16). They also assist in secondary hemostasis by providing a scaffold for assembly of coagulation, and by delivery of various hemostasis proteins (as contained in their storage granules) to the site of injury (1,15,16). These activities are achieved by a sequence of steps that involves platelet adhesion [via various surface receptors and von Willebrand factor (VWF; present within plasma and also stored inside the platelets)], platelet activation (which then causes release of the internal storage granule components) and finally platelet aggregation (in which the platelets aggregate (clump together) to help form the ‘brickwork’ that seals the site of injury. Various plasma proteins, some of which are also included within the platelets, but primarily fibrinogen, help to act as the ‘mortar’ for this brickwork.

Irrespective, the important piece of information to remember in terms of IQC and EQA is that platelets represent small cellular components derived from blood. Although collection of platelets (or of whole blood) is very strait forward, platelets are very easily activated, and once activated can no longer be assessed for their activity or their function (since the activity has already essentially taken place). Platelets can easily become activated by (poor) blood collection (procedures), including use of too small-bore needles, and usage of extended stasis or tourniquets (17). Platelets can also become activated by (poor) blood transport, including excessive agitation (‘shaking’) or pressure (e.g., tube transport systems), delays in transport, and extremes of temperature during transport (both high and low temperatures can adversely affect platelet activity).

In general, PFTs need to be completed within a few hours of (a well-managed) blood collection, so that tests usually need to be performed within the same general location as the blood collection, and after collection of blood by experienced phlebotomists. This is unlike most assays of hemostasis, which can either be performed soon after collection, or else, the plasma can be separated from the centrifuged blood and then frozen for later (even geographically distant) testing. Platelets cannot be frozen for later testing, as freezing platelets destroys them. Platelets cannot be transported over large distances.

IQC and EQA for most tests of hemostasis can typically be achieved using lyophilised or frozen plasma control material. Again, this is simply not possible for PFTs. For most tests of hemostasis, before patient samples can be tested, the process needs to be ‘controlled’ by use of IQC material that can identify whether or not the test processes are working appropriately. The IQC samples must yield test results that are within an acceptable pre-defined range of expected values. Typically, IQC is performed using commercially available (plasma) materials representing several ‘levels’ of the analyte to be controlled, usually meaning a normal sample to control test results around the normal range (or reference interval) and also a ‘pathological’ sample, to control test results above or below the normal range. Sometimes, laboratories utilise a sample that mimics an ‘anticoagulated sample’ to control testing performed around the ‘therapeutic range’ (as an alternative to a ‘pathological’ sample). Such processes can be exemplified, for example, for routine tests of hemostasis, such as prothrombin time (PT), international normalized ratio (INR) and activated partial thromboplastin time (APTT) (18,19), with 2–3 levels of IQC normally applied (20). Indeed, the same set of commercial materials can typically be used for all routine tests, as well as additional tests of hemostasis (e.g., same control set can be used for PT, APTT, thrombin time and fibrinogen).

For PFTs, there are no such commercial controls, simply because as stated, lyophilisation or freezing of platelets, or whole blood, effectively destroys the platelets (21), and there are no commercial sources of stabilised native platelets. Thus, the only way to control PFTs is to collect fresh whole blood or platelets from human donors. To control testing that provides ‘normal’ results, an ostensibly normal individual would need to be collected. This needs to be done each and every time that PFTs are performed—meaning that either a normal individual is punctured every day, or else that the laboratory needs an armamentarium of normal individuals from which to regularly source fresh whole blood/platelets. Neither of these options is ethical or sustainable. Even should such options be available, the arising IQC does not provide control cover for ‘pathological’, ‘therapeutic’ or ‘abnormal’ test results, which for PFTs may involve a multitude of distinct activities depending on the test performed. As a comparative example, for the INR, a ‘therapeutic’ IQC could be a warfarin-like plasma to cover vitamin K antagonist (VKA) therapy, and for the APTT, a ‘therapeutic’ IQC could be a (unfractionated) heparin-like sample (18,19). For PFTs involving LTA and WBA, for example, this might need to include ‘aspirin-treated’ platelets to assess arachidonic acid/cyclooxygenase pathway defects, ‘clopidogrel-treated’ platelets to assess P2Y12 defects and ADP responsiveness, ‘GPIb-denuded’ platelets to assess platelet glycoprotein 1b pathway defects, and so on and so forth. Thus, even if ‘normal’ platelet function testing could be controlled by collection of normal individuals, the results of ‘abnormal’ PFTs cannot be easily controlled, since this would require collection of a wide range of ‘abnormal’ test samples, representing an armamentarium of ‘abnormal’ individuals, or else, the ‘construction’ of abnormal test samples covering a wide variety of potential defects. This is simply not feasible.

EQA for PFTs poses additional challenges. EQA would theoretically require collection of a huge amount of normal blood (to assess/control for normal PFT results) as well as much larger quantities of blood from various abnormal individuals, and/or samples be purpose constructed to reflect a wide variety of potential platelet function abnormalities. This material would then need to be transported to a multitude of EQA participating laboratories, within a few hours (to maintain platelet integrity) and to mitigate adverse effects on platelet function. This is simply a current logistic impossibility.


The (potential) solutions to IQC and EQA for PFTs—focus on the PFA-100/-200

Given the difficulty of IQC and EQA for PFTs (summarised in Table 2) as otherwise ‘classically defined’ for hemostasis tests, there is a need to think ‘outside the box’ and devise alternate strategies to control laboratory performed PFTs and thereby ensure the quality of such testing.

Table 2

Main problems with platelet function tests (PFTs), and performance of internal quality control (IQC) and external quality assessment (EQA)

Platelets are easily activated and can be activated simply by the process of blood collection and/or transport
Once activated, the platelets can no longer be assessed for their activity or ‘function’
Timeliness of PFTs—need to be performed within hours of blood collection
IQC requires collection of fresh whole blood/platelets on regular basis (e.g., daily, if PFTs performed daily)—this carries ethical and sustainable concerns
IQC typically involves assessment of ‘normal’ and ‘pathological/therapeutic’ regions of assay performance. Although the former can be controlled by collection of normal individuals, the latter would require collection of a wide range of ‘abnormal’ test samples, or else, the ‘construction’ of abnormal test samples covering a wide variety of potential defects—this is not feasible
EQA provides even greater challenges. EQA would theoretically require collection of a huge amount of normal blood (to assess/control for normal PFT results) as well as a much larger quantity of blood from abnormal individuals, and/or purpose constructed to reflect a wide variety of potential platelet function abnormalities, and then this blood transported to a multitude of participating laboratories within a few hours of collection and without adverse effect on platelet function. This is simply a current logistic impossibility

The feasibility of alternate approaches to IQC and EQA for PFA-100/-200 testing has been explored by two EQA groups. The College of American Pathologists (CAP) reported on one potential strategy in 2007 (22). This approach utilised an inhibitor of platelet function to generate ‘pathological’ PFT results in the PFA-100 after addition of whole blood from a normal individual, as collected fresh on site by the participating laboratory. Thus, the EQA (CAP in this case) sent the laboratory a ‘wet-challenge’ tube, with this containing a platelet function antagonist, and the EQA participating laboratory then tested normal whole blood collected fresh on site on their PFA-100, either without any manipulation (to generate a normal PFA-100 EQA test sample result) and also repeating the test(s) after addition of this normal whole blood to the ‘wet-challenge’ tube (to generate a ‘pathological’ PFA-100 EQA test sample result). Test results obtained by participant laboratories were then compared with other laboratories in a peer-assessment process. CAP reported this produced a successful EQA for the PFA-100 (22), but inter-laboratory co-efficient of variation (CVs) produced for the wet-challenges were as high as 50%, as compared to about 20% for the normal (non-manipulated test sample), thereby potentially limiting the overall usefulness of the EQA. CAP recently published an update (23) to the original report (22), and the PFA-100 based program remains available more than a decade later in 2019 (https://www.cap.org/). Indeed, this EQA program now also has wet-challenges for ‘Platelet aggregation’ and ‘Helena Plateletworks’, according to the latest CAP catalogue (available at https://www.cap.org/). Moreover, the specific nature of these ‘wet-challenges’, although not identified in the catalogue, is elaborated on in the recent publication (23). It seems that essentially the same approach is used for all the platelet function options—PFA-100, Platelet aggregation, Helena Plateletworks and also PlateletMapping (23). Two specimen challenge tubes are sent with each platelet function survey. One tube contains saline, while the other contains tirofiban, a platelet GPIIb/IIIa inhibitor. If a normal donor is collected, the saline challenge should provide normal results. In contrast, tirofiban would simulate a severe platelet aggregation defect similar to homozygous GPIIb/IIIa deficiency (Glanzmann thrombasthenia). Thus, tirofiban would be expected to produce an abnormal result for all types of platelet function testing. Participating laboratories are instructed to pipette citrated whole blood from a normal donor into each of the challenge tubes and mix gently by inversion 8 to 10 times.

Platelet function testing is then to be performed on these samples according to the standard procedure for each laboratory.

As a summary of their most recent report (23), and using proficiency testing data from 2012–2016, a total of 1,200 laboratories participated for PFA-100, with the coefficient variation (CV) of cartridge closure times for saline being was 22%. The CV for the tirofiban challenge was not reported. Nevertheless, 44,952 of 45,616 survey responses (99%) provided an interpretation, and 42,934 of 44,952 (96%) were correct. This indicated that the wet-challenge process for the PFA-100 worked as expected for the vast majority of participants. For optical platelet aggregation, 190 laboratories participated, and the CV for saline was 17%. Again, the CV for the tirofiban challenge was not reported. Nevertheless, 7,444 of 7,813 survey responses (95%) provided an interpretation, and 7,015 of 7,444 (94%) were correct. This again indicates that this wet-challenge process, this time for optical platelet aggregation, also worked as expected for the vast majority of participants. For PlateletWorks, 60 laboratories participated, and the CV was 3% to 11% (for saline challenge). Of 2,454 survey responses, 2,412 (98%) provided an interpretation, and 1,207 of 1,276 (95%) were correct for adenosine diphosphate (ADP) and 936 of 1,136 (82%) for collagen. For PlateletMapping, 200 laboratories participated; for ADP, 1,128 of 2,697 survey responses (42%) provided an interpretation, but only 927 of 1,128 (82%) were correct. For arachidonic acid, 1,139 of 2,604 survey responses (44%) provided an interpretation and 964 of 1,139 (85%) were correct. Thus, PlateletWorks using collagen and PlateletMapping showed worse interpretive accuracy than the other methods.

As a hemostasis advisor to the Royal College of Pathologists of Australasia (RCPA) hematology quality assurance program (QAP), the author developed an Australasian program for PFA-100 EQA in 2008, and the results of this EQA have since been reported in several publications (24-28). The premise of this EQA is similar to that of CAP, but alternate use of various (propriety) formulations to tirofiban (as used by CAP) has seemingly achieved a much tighter CV than that reported by CAP (22,23). A summary of the anticipated test patterns obtained for the PFA-100/-200 from the perspective of clinical scenarios is provided in Table 3, whereas the corresponding perspective of EQA is shown in Table 4.

Table 3

Expected PFA-100/200 test patterns for different clinical scenariosa

C/Epi C/ADP
Normal Mildly Prolonged Grossly Prolonged (or non-closure)
Normal Normal (mild defectb) Rare event Shouldn’t happen (repeat tests)
Mildly prolonged Aspirin, mild defectb, mildly reduced hematocrit +/− platelet count Mild defectb, mildly reduced hematocrit +/− platelet count Shouldn’t happen (repeat tests)? Severe defectc
Grossly prolonged (or non-closure) Aspirin Moderate to severe defectc, reduced hematocrit +/− platelet count (aspirin) Severe defectc, severely reduced hematocrit +/− platelet count, gross sample hemolysis

a, table summarizes expected PFA-100/-200 test patterns for various clinical scenarios as may be encountered by laboratories undertaking PFA-100/200 testing. b, for example, mild type 1 von Willebrand disease, mild platelet dysfunction. c, for example, type 2A, 2B, 2M, or 3 von Willebrand disease, severe platelet dysfunction.

Table 4

Anticipated PFA-100/200 test patterns for different EQA scenariosa

C/Epi C/ADP
Normal Mildly prolonged Grossly prolonged (or non-closure)
Normal Normal Rare event Shouldn’t happen (repeat tests)
Mildly prolonged Aspirin or mild defect Mild defect Shouldn’t happen (repeat tests)? Severe defect
Grossly prolonged (or non-closure) Aspirin Mild to severe defect Severe defect

a, table summarizes expected PFA-100/-200 interpretations for various test patterns as may be encountered by laboratories undertaking the PFA-100/200 EQA challenge. These potential scenarios are ‘stripped down’ from actual test practice shown in Table 1. For example, the possibilities of low hematocrit and platelet count are ‘ignored’ for pragmatic reasons (essentially, a normal test sample is used in the challenge, albeit post an EQA challenge, and laboratories have difficulties conceptualizing the possibility of low hematocrits and platelet counts in such a setting).

A summary of the EQA material sent out by the RCPA QAP over the past 10 years is identified in Table 5, and the resultant outcomes summarised in Table 6. This EQA provides both positive (‘abnormal’ PFA) and ‘negative’ (‘normal’ PFA) wet challenges, and thus like CAP also provides an EQA wet-challenge for both normal and ‘pathological’ test results. Figure 1 shows data from Table 1 summarised according to challenge type, for main categories of ‘baseline’, ‘no additive challenge’, ‘mild/moderate defect’ challenges (various samples) and ‘severe defect’ challenges (various samples). Inter-laboratory CVs tend to be <20% for baseline test results and no additive ‘wet-challenges’ and are often <15% for ‘positive’ (severe defect’) challenges yielding grossly prolonged PFA test times (Table 6; Figure 1). CVs for ‘mild/moderate defect’ challenges tend to be higher, but are generally <30%, albeit representing quite a heterogeneous group of challenge samples. The EQA has been shown to be effective for both the PFA-100, and the newer PFA-200 model (not yet available in the USA). To date, a total of 47 separate EQA challenges have been distributed and undertaken by participants (Table 5), with results published for most challenges (24-28). A summary of data for baseline PFA test times, the negative ‘wet-challenges’, and some positive ‘wet-challenges’ is shown in Figure 2.

Table 5

Summary of external quality assessment (EQA) trials undertaken by participants of the RCPAQAP to date

Year Type of survey Number of participants Number of samples Sample types/scenarios that samples designed to mimic
2008 Trial 26 5 Normal baseline CTs; mild defect; severe defect
2009 Trial 26 6 Normal baseline CTs; aspirin defect; mild defect; severe defect
2010 Formal EQA module 47 4 Normal baseline CTs; severe defect
2011 Formal EQA module 47 4 Normal baseline CTs; moderate defect; severe defect
2012 Formal EQA module 49 4 Normal baseline CTs; moderate defect; severe defect
2013 Formal EQA module 50 4 Normal baseline CTs; aspirin defect; severe defect
2014 Formal EQA module 53 4 Normal baseline CTs; severe defect
2015 Formal EQA module 59 4 Normal baseline CTs; moderate defect; severe defect
2016 Formal EQA module 58 4 Normal baseline CTs; moderate defect; severe defect
2017 Formal EQA module 65 4 Normal baseline CTs; moderate defect; severe defect
2018 Formal EQA module 73 4 Normal baseline CTs; severe defect
Total 47

RCPAQAP, Royal College of Pathologists of Australasia Quality Assurance Program.

Table 6

Summary of findings of wet challenges undertaken by participants of the RCPAQAP PFA-100/-200 test module to date

Year and wet challenge sample identitya Scenario that sample designed to mimic Target PFA-100/200 CTsb Median CTs CVs (%)
C/ADP (s) C/Epi (s) C/ADP (s) C/Epi (s) C/ADP C/Epi
2008-baseline Normal baseline CTs Normal Normal 81 111 14.3 18.5
2008-2 Mild defect ~150–200 ~200 141 170 27.8 17.3
2008-3 Severe defect >250 >250 301 301 13.3 15.1
2008-4 Mild defect ~150–200 ~200 170 210 22.0 23.7
2008-5 Severe defect >250 >250 301 301 14.8 14.6
2009-Baseline ‘a’ Normal baseline CTs Normal Normal 97 125 18.0 19.1
2009-1a Aspirin effect Normal >250 95 301 16.4 27.1
2009-2a Severe defect >250 >250 296 301 14.7 11.6
2009-3a Severe defect >250 >250 301 301 4.4 12.6
2009-Baseline ‘b’ Normal baseline CTs Normal Normal 93 107 15.4 18.8
2009-1b Mild defect ~150–200 ~200 167 175 29.5 29.5
2009-2b Severe defect >250 >250 291 301 10.9 13.9
2009-3b Severe defect >250 >250 301 301 3.9 0.6
2010 Dispatch 1 baseline Normal baseline CTs Normal Normal 84 119 15.4 14.1
2010 PF10-03a Severe defect >250 >250 301 301 19.4 12.0
2010 PF10-03b Normal (no additive tube) Normal Normal 100 130 18.4 15.2
2010 Dispatch 2 baseline Normal baseline CTs Normal Normal 86 115 15.1 16.6
2010 PF10-08a Severe defect >250 >250 301 301 4.6 12.4
2010 PF10-08b Normal (no additive tube) Normal Normal 99 130 18.9 13.5
2011 Dispatch 1 baseline Normal baseline CTs Normal Normal 87 118 15.8 20.5
2011 PF11-03a Normal (no additive tube) Normal Normal 94 130 17.9 18.1
2011 PF11-03b Severe defect >250 >250 301 301 17.4 14.6
2011 Dispatch 2 baseline Normal baseline CTs Normal Normal 89 130 14.2 14.1
2011 PF11-08a Moderate/severe defect >200 >200 223 301 23.0 12.0
2011 PF11-08b Severe defect >250 >250 301 301 7.3 5.1
2012 Dispatch 1 baseline Normal baseline CTs Normal Normal 84 119 14.2 15.6
2012 PF12-03a Normal (no additive tube) Normal Normal 96 132 19.9 17.3
2012 PF12-03b Moderate/severe defect >200 >200 214 301 29.5 16.6
2012 Dispatch 2 baseline Normal baseline CTs Normal Normal 83 116 15.7 15.0
2012 PF12-08a Moderate/severe defect >200 >200 214 301 29.4 16.4
2012 PF12-08b Severe defect >250 >250 301 301 13.5 15.2
2013 Dispatch 1 baseline Normal Baseline CTs Normal Normal 88 118 17.5 19.0
2013 PF13-03a Severe defect >250 >250 301 301 15.9 11.4
2013 PF13-03b Aspirin effectc Normal >250c 95 134c 24.6 38.4c
2013 Dispatch 2 baseline Normal baseline CTs Normal Normal 86 120 15.4 15.9
2013 PF13-08a Normal (no additive tube) Normal Normal 93 130 25.7 23.1
2013 PF13-08b Severe defect >250 >250 301 301 15.5 8.2
2014 Dispatch 1 baseline Normal baseline CTs Normal Normal 84 116 15.9 16.1
2014 PF14-03a Normal (no additive tube) Normal Normal 95 123 20.0 22.6
2014 PF14-03b Severe defect >250 >250 301 301 17.1 14.9
2014 Dispatch 2 baseline Normal baseline CTs Normal Normal 83 130 15.1 15.3
2014 PF14-08a Severe defect >250 >250 301 301 19.1 10.2
2014 PF14-08b Moderate/severe defect >200 >200 301 301 9.8 10.1
2015 Dispatch 1 Baseline Normal baseline CTs Normal Normal 88 119 17.5 18.0
2015 PF15-03a Moderate/severe defect >200 >200 301 301 16.9 5.4
2015 PF15-03b Normal (no additive tube) Normal Normal 98 132 26.2 29.8
2015 Dispatch 2 baseline Normal baseline CTs Normal Normal 87 117 16.6 15.5
2015 PF15-08a Severe defect >250 >250 301 301 16.2 13.2
2015 PF15-08b Normal (no additive tube) Normal Normal 97 127 22.1 24.3
2016 Dispatch 1 baseline Normal baseline CTs Normal Normal 83 116 12.6 15.7
2016 PF16-03a Severe defect >250 >250 301 301 9.1 12.8
2016 PF16-03b Normal (no additive tube) Normal Normal 94 122 18.8 25.2
2016 Dispatch 2 baseline Normal baseline CTs Normal Normal 86 113 17.9 21.6
2016 PF16-08a Normal (no additive tube) Normal Normal 97 129 22.1 24.5
2016 PF16-08b Moderate/severe defect >200 >200 301 301 23.0 13.1
2017 Dispatch 1 baseline Normal baseline CTs Normal Normal 84 116.5 16.7 17.0
2017 PF17-03a Moderate/severe defect >200 >200 301 301 15.7 11.8
2017 PF17-03b Normal (no additive tube) Normal Normal 95 125 19.2 29.1
2017 Dispatch 2 baseline Normal Baseline CTs Normal Normal 85 118 15.2 16.7
2017 PF17-08a Normal (no additive tube) Normal Normal 97 125 18.9 18.2
2017 PF17-08b Severe defect >250 >250 301 301 15.6 13.3
2018 - Dispatch 1 baseline Normal baseline CTs Normal Normal 82 117 19.3 16.9
2018 PF18-03a Severe defect >250 >250 301 301 5.9 8.5
2018 PF18-03b Normal (no additive tube) Normal Normal 94 121 18.2 18.6
2018 Dispatch 2 baseline Normal baseline CTs Normal Normal 84 118 19.5 19.7
2018 PF18-08a Normal (no additive tube) Normal Normal 95 125 19.4 23.1
2018 PF18-08b Severe defect >250 >250 301 301 8.0 7.6

a, five challenge samples dispatched in 2008; data for one sample showing stability issues omitted. Six challenge samples dispatched to the same laboratories in the trial 2009 exercise, but testing was split into two sets of three samples. The formal PFA-100 external quality assessment (EQA) module began in 2010, where two samples were dispatched to 42 participants in March and another two samples to 47 participants in August. A similar dispatch process has been used thereafter. Some similarly or identically formulated challenge samples were dispatched in different exercises to help assess reproducibility of the system. ‘Baseline’ data represents data within each exercise using native whole blood prior to test challenges. b, ‘normal’ means CTs within the normal reference range. c, challenge PF13-03b was designed as an aspirin-challenge, and despite acceptable homogeneity testing, showed unacceptable stability findings, and lead to a failed EQA challenge. All other challenges were essentially deemed to be successful challenges. Similar sample sets are identified by scenario (e.g., all challenges that represent no additive, or all challenges that mimic severe defect. Identical sample sets are those that comprise the same challenge material sent in different surveys; viz: 2008-3 & 2009-3a; 2008-4 & 2009-1b; 2008-5 & 2009-3b; PF10-08a & PF11-08b & PF12-08b; PF14-03b & PF14-08a; PF14-08b & PF15-03a; PF15-08a & PF16-03a; PF17-08b & PF17-03a. RCPAQAP, Royal College of Pathologists of Australasia Quality Assurance Program; CTs, closure times; CVs, coefficient of variation; C/ADP, collagen/ADP; C/Epi, collagen/epinephrine.

Figure 1 Summary of data findings from Table 6. Data summarised according to challenge type, for main categories of ‘baseline’ (B), ‘no additive challenge’ (NA), ‘mild/moderate defect’ challenges (M; various samples) and ‘severe defect’ challenges (S; various samples). (A) Median values for PFA closure times (CTs) in seconds (s) for collagen/ADP (C/ADP) and collagen/epinephrine (C/Epi) cartridges. The dotted horizontal lines represent the upper limit of normal using manufacturer reported ranges. (B) Inter-laboratory coefficient of variation (CVs). The dotted horizontal lines represent 10%, 20% and 30%.
Figure 2 Summary of external quality assessment (EQA) data for PFA-100/-200 from Royal College of Pathologists of Australasia (RCPA) hematology Quality Assurance Program (QAP), for years 2008–2017 inclusive. Data shown as box plots, with the bars representing the 10th–90th percentiles, and the box representing the 25th–75th percentiles. Left y-axis collagen/ADP (C/ADP) cartridge closure time (CT) in seconds. Right y-axis collagen/epinephrine (C/Epi) cartridge closure time (CT) in seconds. X-axis identifies the EQA challenge. The dotted horizontal lines represent the upper limit of normal using manufacturer reported ranges. (A) Baseline (pre-wet-challenge) data; (B) ‘negative’ (no additive) wet challenges; (C) ‘positive’ (additive) wet challenges.

The PFA EQA has also been shown to be potentially useful in the setting of IQC (26-28). For the PFA-100/-200, the manufacturer only suggests performance of a normal test sample (normal fresh donor) with each change in lot of test cartridges, or after any major instrument maintenance, as IQC. The EQA ‘wet-challenges’ may be therefore potentially utilised to also provide some reassurance of PFA functionality around the ‘abnormal’ test region, and test data can be depicted in Levy-Jennings graphs (25-28) (samples shown in Figure 3).

Figure 3 Levey-Jennings-like plots of an internal quality control (IQC) like process using some external quality assessment (EQA) challenge samples. (A,B) Normal baseline whole blood closure time (CT) values can act as ‘normal IQC’ and sequential test data from EQA samples PF12-08b (A) and PF15-08a (B) can act as ‘pathological IQC’. Data reported as CTs in seconds (s; y-axis in each figure), for a theoretical timeline of 12 months. Here, IQC limits for the ‘normal QC’ sample (= baseline whole blood CTs) would be values below the normal/abnormal cut-off value (manufacturer values used in this example). For PFA-12-08b and PF15-08a, representing ‘pathological’ IQC samples, the IQC limit could be assigned as a value above a predefined cut-off (e.g., 200s as used in this example). (C,D) Data from Figures A and B, but now expressed in terms of x-fold of baseline. In this potential IQC scenario, the IQC limits might be expressed by predefined limits of test data (for example above 2x baseline value).

Other EQA solutions for PFTs

Another EQA provider, ECAT (External Quality Control of Diagnostic Assays and Tests; http://www.ecat.nl/) in partnership with NASCOLA (North American Specialized Coagulation Laboratory Association; https://www.nascola.com/) offer a variety of ‘electronic’ surveys to support laboratories involved in PFTs. These comprise: (I) Post Analytical Platelet Function EQA (electronic survey); (II) Platelet Dense Granule exercise (electronic survey); (III) case studies on bleeding disorders (distribution separately from the regular surveys); (IV) pre- and post-analytical electronic surveys in haemostasis. Some publication around these exercises are available for the interested reader (29-32).


Conclusions

IQC and EQA for PFTs remains challenging, but some support is available from a variety of sources, especially in relation to EQA. Wet-challenges are so far limited to two EQA providers, RCPAQAP and CAP, although ECAT expects to also offer the same challenges as available to RCPAQAP participants in the near future. In terms of EQA for PFTs, the main alternative to ‘wet-challenges’ comes in the form of educational support, for example by electronic surveys, and typically ‘post-analytical’ (meaning that results of PFTs are provided the EQA to participants for their interpretation – and then these interpretations are then relayed to the EQA for review and assessment/peer comparison/reporting). Irrespective, ‘thinking outside the box’ is a critical requirement when contemplating IQC and EQA for PFTs, since standard hemostasis approaches simply remain infeasible (Table 2).


Acknowledgments

This review includes some data derived as a result of an ongoing collaboration with the RCPAQAP. This data is presented in summary form, and/or may derive from prior studies in which the RCPAQAP and its staff contributed, either by provision of raw data, or by prior data analysis, or by other logistical support. This EQA program is therefore acknowledged as facilitating some of the work described in this paper. Several individuals within the RCPA can also be personally acknowledged, in particular Roslyn Bonar, who has recently retired from the workforce to concentrate on life and family.


Footnote

Provenance and Peer Review: This article was commissioned by the editorial office, Annals of Blood for the series “External Quality Assurance”. The article has undergone external peer review.

Conflicts of Interest: The author has completed the ICMJE uniform disclosure form (available at: http://dx.doi.org/10.21037/aob.2019.01.02). The series “External Quality Assurance” was commissioned by the editorial office without any funding or sponsorship. EJF served as an unpaid Guest Editor of the series. The author has no other conflicts of interest to declare.

Ethical Statement: The author is accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Disclaimer: The opinions expressed in this review are those of the author, and not necessarily those of NSW Health Pathology or of the RCPAQAP. This manuscript reflects findings from a single EQA provider, the RCPAQAP, and the opinions of other EQA organizations may differ.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Gremmel T, Frelinger AL 3rd, Michelson AD. Platelet Physiology. Semin Thromb Hemost 2016;42:191-204. [Crossref] [PubMed]
  2. Hvas AM. Platelet Function in Thrombosis and Hemostasis. Semin Thromb Hemost 2016;42:183-4. [Crossref] [PubMed]
  3. Gresele P, Bury L, Falcinelli E. Inherited Platelet Function Disorders: Algorithms for Phenotypic and Genetic Investigation. Semin Thromb Hemost 2016;42:292-305. [Crossref] [PubMed]
  4. Harrison P, Mackie I, Mumford A, et al. British Committee for Standards in Haematology. Guidelines for the laboratory investigation of heritable disorders of platelet function. Br J Haematol 2011;155:30-44. [Crossref] [PubMed]
  5. Favaloro EJ, Lippi G, Franchini M. Contemporary platelet function testing. Clin Chem Lab Med 2010;48:579-98. [Crossref] [PubMed]
  6. Lordkipanidzé M. Platelet Function Tests. Semin Thromb Hemost 2016;42:258-67. [Crossref] [PubMed]
  7. Lassila R. Platelet Function Tests in Bleeding Disorders. Semin Thromb Hemost 2016;42:185-90. [Crossref] [PubMed]
  8. Fritsma GA, McGlasson DL. Whole Blood Platelet Aggregometry. Methods Mol Biol 2017;1646:333-47. [Crossref] [PubMed]
  9. Hvas AM, Favaloro EJ. Platelet Function Analyzed by Light Transmission Aggregometry. Methods Mol Biol 2017;1646:321-31. [Crossref] [PubMed]
  10. Favaloro EJ. Clinical utility of closure times using the Platelet Function Analyzer (PFA)-100/200. Am J Hematol 2017;92:398-404. [Crossref] [PubMed]
  11. Ramström S, Södergren AL, Tynngård N, Lindahl TL. Platelet Function Determined by Flow Cytometry: New Perspectives? Semin Thromb Hemost 2016;42:268-81. [Crossref] [PubMed]
  12. Cai H, Mullier F, Frotscher B, et al. Usefulness of Flow Cytometric Mepacrine Uptake/Release Combined with CD63 Assay in Diagnosis of Patients with Suspected Platelet Dense Granule Disorder. Semin Thromb Hemost 2016;42:282-91. [Crossref] [PubMed]
  13. Pasalic L. Assessment of Platelet Function in Whole Blood by Flow Cytometry. Methods Mol Biol 2017;1646:349-67. [Crossref] [PubMed]
  14. Gross L, Aradi D, Sibbing D. Platelet Function Testing in Patients on Antiplatelet Medications. Semin Thromb Hemost 2016;42:306-20. [Crossref] [PubMed]
  15. Bonar RA, Lippi G, Favaloro EJ. Overview of Hemostasis and Thrombosis and Contribution of Laboratory Testing to Diagnosis and Management of Hemostasis and Thrombosis Disorders. Methods Mol Biol 2017;1646:3-27. [Crossref] [PubMed]
  16. Lippi G, Favaloro EJ. Laboratory hemostasis: from biology to the bench. Clin Chem Lab Med 2018;56:1035-45. [Crossref] [PubMed]
  17. Lippi G, Favaloro EJ. Preanalytical Issues in Hemostasis and Thrombosis Testing. Methods Mol Biol 2017;1646:29-42. [Crossref] [PubMed]
  18. Favaloro EJ. Optimizing the Verification of Mean Normal Prothrombin Time (MNPT) and International Sensitivity Index (ISI) for Accurate Conversion of Prothrombin Time (PT) to International Normalized Ratio (INR). Methods Mol Biol 2017;1646:59-74. [Crossref] [PubMed]
  19. Kershaw G. Performance of Activated Partial Thromboplastin Time (APTT): Determining Reagent Sensitivity to Factor Deficiencies, Heparin, and Lupus Anticoagulants. Methods Mol Biol 2017;1646:75-83. [Crossref] [PubMed]
  20. Bonar R, Favaloro EJ, Adcock D. Quality in coagulation and haemostasis testing. Biochemia Medica 2010;20:184-99. [Crossref]
  21. Favaloro EJ. Internal Quality Control and External Quality Assurance of Platelet Function Tests. Semin Thromb Hemost 2009;35:139-49. [Crossref] [PubMed]
  22. Cunningham MT, Brandt JT, Chandler WL, et al. Quality assurance in hemostasis: the perspective from the College of American Pathologists proficiency testing program. Semin Thromb Hemost 2007;33:250-8. [Crossref] [PubMed]
  23. Chandler WL, Brown AF, Chen D, et al. External Quality Assurance of Platelet Function Assays Results of the College of American Pathologists Proficiency Testing Program. Arch Pathol Lab Med 2018; [Crossref]
  24. Favaloro EJ, Bonar R. External quality assurance for the PFA-100. J Thromb Haemost 2011;9:878-80. [Crossref] [PubMed]
  25. Favaloro EJ, Bonar R. Proficiency testing/External quality assurance for the PFA-100. Clin Chem Lab Med 2012;50:1393-401. [Crossref] [PubMed]
  26. Favaloro EJ. Time for a conceptual shift in assessment of internal quality control for whole blood or cell based testing systems? An evaluation using platelet function and the PFA-100 as case example. Clin Chem Lab Med 2013;51:767-74. [Crossref] [PubMed]
  27. Favaloro EJ, Bonar R. External Quality Assessment/Proficiency Testing and Internal Quality Control for the PFA-100 and PFA-200: An Update. Semin Thromb Hemost 2014;40:239-53. [Crossref] [PubMed]
  28. Favaloro EJ, Bonar R. An update on quality control for the PFA-100/PFA-200. Platelets 2018;29:622-7. [Crossref] [PubMed]
  29. Hayward CP, Moffat KA, Spitzer E, et al. NASCOLA Working Group on Platelet Dense Granule Deficiency. Results of an external proficiency testing exercise on platelet dense-granule deficiency testing by whole mount electron microscopy. Am J Clin Pathol 2009;131:671-5. [Crossref] [PubMed]
  30. Hayward CP, Moffat KA, Raby A, et al. Development of North American Consensus Guidelines for Medical Laboratories That Perform and Interpret Platelet Function Testing Using Light Transmission Aggregometry. Am J Clin Pathol 2010;134:955-63. [Crossref] [PubMed]
  31. Hayward CP, Moffat KA, Plumhoff E, et al. External quality assessment of platelet disorder investigations: results of international surveys on diagnostic tests for dense granule deficiency and platelet aggregometry interpretation. Semin Thromb Hemost 2012;38:622-31. [Crossref] [PubMed]
  32. Chen D, Uhl CB, Bryant SC, et al. Diagnostic Laboratory Standardization and Validation of Platelet Transmission Electron Microscopy. Platelets 2018;29:574-82. [Crossref] [PubMed]
doi: 10.21037/aob.2019.01.02
Cite this article as: Favaloro EJ. Novel approaches to quality control and external quality assessment for platelet function testing with a focus on the platelet function analyser (PFA-100 and PFA-200). Ann Blood 2019;4:3.